BigBite Optics et cetera #17
Results of the Th/Ph optics calibration:
Run #3488 : Hydrogen Elastic
Run #3489 : Carbon Quasielastic
Run #3490 : Carbon Quasielastic
Run #3491 : Carbon Quasielastic
Run #3493 : Carbon Quasielastic - BigBite Field off
This run was used only for the calibration of holes on top.
Run #3495 : Carbon Quasielastic - BigBite Field off
This run was used only for the calibration of holes on top.
Run #3496 : Empty Cell run- No sieve
Th - Offset correction
We still have problems with the offsets in both, vertical and horizontal axis.
The results for the elastic hydrogen run are not consistent with the left-arm results:
Where do these offsets come from?
Hydrogen Elastic run - HRSL vs. BigBite
From plots below we can see that elastic protons which correspond to scattered electrons with TgTh = 0,
do not go through the center line of holes in BB Sieve, but hit the collimator somewhere between the
central line and one line below:
Short schematic explanation: PDF with schematics.
Results after a correction?
Our best survey from 2 years ago : Survey.
My Change in the HRS-L Optics matrix : The Matrix.
Last modified: 07/28/10